Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K172305

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-10-12

    (73 days)

    Product Code
    Regulation Number
    870.3535
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CS100/CS100i and CS300 Intra-Aortic Balloon Pumps are indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure.

    Device Description

    The CS100/CS100i and CS300 Intra-Aortic Balloon Pumps (IABPs) are cardiac assist devices. The IABP supports the heart's left ventricle by increasing coronary perfusion and reducing left ventricular work. Coronary perfusion is increased by augmenting blood pressure during the diastolic phase of the cardiac cycle. This increase in aortic pressure promotes more blood flow through the coronary arteries. Left ventricular work is reduced by decreasing aortic end-diastolic pressure and reducing resistance to ventricular ejection, resulting in a decrease in blood pressure during the systolic phase of the cardiac cycle. These beneficial effects are caused by the inflation and deflation of an intra-aortic balloon (IAB) in the patient's descending aorta. The balloon's inflation and deflation must be properly synchronized with the cardiac cycle. IAB inflation is initiated at the onset of diastole at the dicrotic notch and remains inflated through diastole. The IAB is then deflated at, or just prior to, the onset of systole and the balloon remains deflated throughout systole. Hence, the therapy is also referred to as counterpulsation.

    AI/ML Overview

    The provided text is a 510(k) summary for the CS100/CS100i and CS300 Intra-Aortic Balloon Pumps. It states that the device is substantially equivalent to a predicate device and includes information about modifications made to the device. However, it explicitly states that there was no clinical evaluation of the modified device.

    Therefore, I cannot provide the requested information regarding acceptance criteria, device performance, sample sizes, expert qualifications, and ground truth, as these details are typically derived from clinical studies or extensive performance testing, which were not conducted for the modified device described in this document.

    The document only notes that:

    • The changes consist of a revised blood detection software function.
    • "Performance data" claims that "The results of the tests conducted demonstrate that the modified CS100/CS100i and CS300 Intra-Aortic Balloon Pumps perform as intended."
    • "Datascope Corp. development process required that the following activities be completed...: Requirements specification review, Performance testing, Design verification and validation."

    Without a clinical study, there are no specific acceptance criteria for performance metrics that would typically involve human-in-the-loop performance or comparison against a gold standard established by experts.

    In summary, based on the provided text, the device's substantial equivalence was determined without a clinical study, meaning the details requested (acceptance criteria, performance, sample sizes, expert qualifications, ground truth, MRMC study, standalone performance) are not available within this document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1